IT series medical isolation transformers are designed according to IEC 60364-7-710 standards for powering single-phase IT systems and are manufactured according to IEC 61558-2-15 standards. Medical isolation transformers are compatible with IT systems in medical locations.
IT series medical isolation transformers, with galvanically isolated windings, offer temperature monitoring thanks to PTC thermistors placed in the windings. IT series medical isolation transformers, designed at 6 different power ratings between 3.15 and 10 kVA, have high overload capacities.
Isolation transformers are designed to reduce the effect of electrical faults. Due to their electrostatic protection, isolation transformers are used to power operating rooms, laboratory instruments, intensive care units, and other similar sensitive devices.
Power Range - 3.15 ... 10 kVA
Primary Voltage - 230 Vac
Secondary Voltage - 230/115 Vac
Frequency - 50/60 Hz
Windings - Al or Cu
Inrush Current (Ie) - < 12xIn
Leakage Current - < 0.5 mA
No-load Input Current (Io) - < 3%
No-load Output Current (Io) - < 236 Vac
Short-Circuit Voltage (Uk) - < 3%
Max Ambient Temp. - 40°C
Protection Class - IP 00 (IP 21 optional)
Isolation Class - F (155 °C) or B (120 °C)
Specialized electrical installations of any complexity in Chisinau and throughout Moldova